Elinvo (Electronics, Informatics, and Vocational Education)

Keywords

Kstethoscope modification, heart sound, Matlab GUI

Document Type

Article

Abstract

Heart sounds are very important for the initial diagnosis of a disease. This article aims to modify the stethoscope's medical device by adding a Low Pass Filter (LPF) and a condenser sensor so that data can be recorded in a digital device. The stages of tool development begin with analysis, design, development and testing. At the analysis stage, it was stated that a conventional medical stethoscope can only be done by a medical expert directly, where data cannot be stored for detailed analysis purposes, the device is not equipped with LPF which can limit the frequency in a certain range, monitoring data can displayed in the Matlab GUI. The results of testing tools through functional tests indicate that the addition of LPF can reduce noise when retrieving heart sound data and data can be monitored through the Matlab GUI display. The results of the development of this tool are expected to improve accuracy in diagnosing heart sound abnormalities through monitoring data on the Matlab GUI display.

First Page

106

Last Page

112

Page Range

106-112

Issue

1

Volume

3

Digital Object Identifier (DOI)

10.21831/elinvo.v3i1.20892

Source

https://journal.uny.ac.id/index.php/elinvo/article/view/20892

References

[1] Depkes, (2018,8,13), Penyakit Jantung Penyebab-Kematian, http://www.depkes.go.id/article/view/17073100005/

[2] Utami P., & Fajar, M. "Pengembangan Trainer Indikator Denyut Jantung". Elinvo, 1(1). 2015.

[3] Ira Puspasari,dkk, "Analisis Ekstraksi Ciri Pada Suara Jantung Diastolik dengan menggunakan Wavelet Transform dan Wigner Ville Distribution". NY

[4] NN. "Buku acuan Peserta CSL 2 Seri 3 Pemeriksaan Fisis Jantung". Fak. Kedokteran UNHAS. NY

[5] James N, et al. "Quantitative Analysis of Heart Sounds and Systolic Heart Murmurs Using Wavelet Transform and AR Modeling". IEEE. 2010

[6] Kemalasari, dkk, "Deteksi Kelainan Parenkim Paru Berdasarkan Power Spectra Density Suara Paru Dengan Metode Welch", EEPIS, 2011

[7] Proakis,Monolakis."Digital Signal Processing 4th Edition". Prentice Hall, NY

[8] NN. "Digital Sthetoscope Implementation on the TMS320C5515 DSP Medical Development Kit (MDK). Texas Instrumen, 2010

Included in

Biomedical Commons

Share

COinS